Pintek PS-608 Analog Oscilloscope with Cursor Readout 60MHz

by Pintek
DISCONTINUED
Pintek PS-608 Analog Oscilloscope with Cursor Readout 60MHz
60MHz Bandwidth
CURSOR READOUT FUNCTION
1mV/DIV SENSITIVITY
10nS/DIV RESLUTION
CE, TUV GS. UL, CUL, IEC101

Specifications

CRT
Type : Rectangular with integral graticule.
Display Area : 8 x 10 div ( 1 div = 1 cm )
Accelerating Voltage : 12kV.
Phosphor : P31.

VERTICAL AMPLIFIERS ( CH 1 and CH 2 )
Sensitivity :
5 mV / div to 5 V / div ( X1 mode ).
1 mV / div to 1 V / div ( X5 mode ).
1-2-5 sequence, 10 step attenuator.
Vernier control provides fully adjustable attenuation between steps.
Accuracy : +/- 3 % ( X1 mode );+/- 5 % ( X5 mode ).
Input Resistance : 1 M Omh +/- 2 %.
Input Capacitance : 25 pF +/- 10 pF.
Frequency Response :
5 mV to 5 V / div ( CAL ) : DC to 60 MHz (-3 dB).
1 mV / div (CAL) : DC to 20 MHz (-3 dB).
Rise Time : Approximately 5.8 ns.
Operating Modes -
CH1 : CH 1, single trace.
CH2 : CH 2, single trace.
ALT : Dual trace, alternating.
CHOP : dual trace, chopped.
ADD : algebraic sum of CH 1 + CH 2.
Polarity Reversal (Invert) : CH 2 only.
Maximum Input Voltage : 400 V dc + ac peak.
Maximum Undistorted Amplitude :
DC-to-60 MHz: 4 divisions.
DC-to-40 MHz: 8 divisions.

HORIZONTAL AMPLIFIER ( Input through channel 1 input )
X-Y mode : switch selectable using X-Y switch.
CH 1: X axis.
CH 2: Y axis.
Sensitivity : 1mv/div to 5v/div
Accuracy -
Y-Axis : +/- 3%.
Y-Axis : +/- 6%.
Input Impedance : 1M Omh +/- 2% //25PF
Frequency Response : DC to 1 MHz typical ( -3 dB ).
X-Y Phase Difference : Approximately 3o at 50 kHz.
Maximum Input Voltage : 400v dc+ac pack

SWEEP SYSTEM
Sweep Speed:0.1 us/div to 0.5 s/div in 1-2-5 sequence, 21 steps.
Vernier control provides fully adjustable sweep time between steps.
Accuracy : +/- 3 %.
Sweep Magnification : 10X, +/- 10%
Hold off : Continuously adjustable from NORM to 5 times normal.

TRIGGERING
Trigger Modes : AUTO ( free run ) or NORM.
Trigger Source : CH1, CH2, ALT, EXT, LINE.
Maximum External Trigger Voltage : 300 V dx + ac peak.
Trigger Coupling -
AUTO ( AC ) : 30 Hz to 80 MHz.
TV-H ( AC ) : Used for triggering from horizontal sync pulses.
TV-V ( DC ) : Used for triggering from vertical sync pulses.

READOUT FUNCTION-Panel setting displays
Vertical axis ( CH1, CH2 ) : V/DIV,UNCAL, MAG ( Converted Value)
Horizontal axis : S/DIV, UNCAL, MAG ( Converted Value)
- Cursor readout function
Voltage difference V : to REF
Time difference T : to REF
Frequency difference 1/ T : 1/1 to REF 1
NOTE : When the X-Y mode, V is changed to X and T to Y.
-Cursor measurement resolution : 1/25 DIV
-Effective cursor range from center graticule
Vertical : within +/- 3 DIV
Horizontal : within +/- 4 DIV

OTHER SPECIFICATIONS
Calibrating Voltage : 1 kHz ( nominal ) Positive Square Wave, 0.5 V p-p ( +/- 3% )
Trace Rotation : Electrical, front panel adjustable.
Power Requirements : 115/ 230 VAC +/- 10%, 50/60Hz , approximately 60 W.
Dimensions : 324 x 132 x 398 mm (12.8 x 5.2 x 15.6 in).
Weight : Approximately 8.5 kg (18.7 1bs).

Downloads

In the Box

2 x Pintek CP-260 Test Probes
Power Cord
Operation Manual

Reviews